2 default setting: adjustment with internal weight – KERN & SOHN ABT 220-5DM User Manual

Page 25

Advertising
background image

ABT-BA-e-1312

25

8.3.2 Default setting: Adjustment with internal weight

Condition: Function „i tCAL“ activated (see ch. 8.3.1)

Press the [CAL] key The display shows “i-CAL”

Press the [TARE] key. The display changes automatically in this

order: “CAL 2”, “CAL 1”, “CAL 0”, and “CAL End”. After success-

ful adjustment the balance automatically returns to weighing

mode.

In case of an adjustment error (e.g. objects on the weighing

plate) the display will show an error message, repeat adjustment.

8.3.3 Default setting: Adjustment test with internal weight
During adjustment tests the balance automatically compares the saved value of the

adjustment weight with the actual value. This is only a check, i.e. no values are

changed.


Condition: Function „i tESt“ activated (see ch. 8.3.1)

Press the [CAL] key The display shows “i-tESt”

Press the [TARE] key. The display changes automatically in this

order: “tESt 2”, “tESt 1”, “tESt 0”

After this, the difference to the previous adjustment will be dis-

played for several seconds.

After "tESt End" was displayed, the balance will automatically

return to weighing mode.

Advertising
This manual is related to the following products: