2 standstill display – KERN & SOHN ABT 220-5DM User Manual

Page 49

Advertising
background image

ABT-BA-e-1312

49

13.2 Standstill display

If the display shows the stability display ( ) the balance is in a stable status. The

condition for judging stability is user-selectable. There are three levels to select: 1

count, 5 counts and 10 counts. The default setting is 1 count.

The setting of the standstill display can be checked on the setting control display

(see ch.12.6).
Setting the standstill display

Repeatedly press the [CAL] key until "FUnC.SEL" appears.

Press [TARE] key

(Example)

Repeatedly press the [CAL] key until the current "b And:**" set-

ting appears.

Press [TARE] key

You can use the [CAL] key to select between the following set-

tings:

"b-1"

1 count

"b-1"

5 count

"b-1"

10 count

The current setting is marked by the stability symbol ( ).

Confirm your selection by pressing the [TARE] key

To exit the function, press the [ON/OFF] key.

Brief actuation of ON/OFF key:

Back to previous menu.

Long actuation of ON/OFF key:

Back to weighing mode.