2 percent determination – KERN & SOHN ABT 220-5DM User Manual

Page 55

Advertising
background image

ABT-BA-e-1312

55


Remove reference weight.

Now you can fill the items to be counted into the container.

The respective quantity is shown in the display.

Repeated pressing of the [UNIT] key changes the display value,

e. g.: [g] [%] [Pcs] [ct]


Note:

Error message "Err 20" indicates that the weight for the reference piece number is

too low.

14.2 Percent determination
Percent weighing allows to display weight in percent, in relation to a reference

weight. The displayed weighing value is stored as a standard percent value

(default setting: 100%).


Condition:

Function "U- %" is activated (ch. 11.3)

(When PCS is used for the

first time)

Repeatedly press the [UNIT] key until the display is in percent-

age determination mode. The display symbol % pops up.

If you are using a weighing container use the [TARE] key for

taring.

Put on reference weight (=100%)

(minimum weight: Readability d x 100)

When standstill control ( ) is complete, press the [CAL] key.

The reference weight is saved.

Remove reference weight.

From now on the placed weight will be shown as %.

Repeated pressing of the [UNIT] key changes the display value,

e. g.: [g] [%] [Pcs] [ct]

Advertising