ABT-BA-e-1312

56

14.3 Density determination

Density determination with the help of the flush-mounted platform facility is described

in the following.

Density determination becomes even easier when an optional set for density deter-

mination is applied. For further information please refer to the operating instructions

enclosed with the set for density determination.

1. Remove the below-weigh hook cover on the floor of the balance, after remov-

ing the two fixing screws.

2. Hang the user-prepared hanging pan from the below-weigh hook and sink that

hanging pan into the sample liquid in the tank.

14.3.1 Density determination of solids

Condition

: Function ",d" (density of solids) is activated. See Chapter 11.3.

Press the [CAL] key repeatedly until "SettinG" appears.

Press the

[TARE]

key.

Repeatedly actuate the

[CAL]

key until "LSG SEt" appears

(Example)

Press the [TARE] key. The currently set density for the liquid to be

measured appears. In the upper part of the display panel, the

symbol and the # symbol appear in order to indicate numer-

ical input status. The leftmost digit blinks.

(Example)

Enter density for your liquid to be measured. When the [UNIT] key

is pressed, the numerical of the blinking digit increases by 1 at a

time. You can determine the value of the flashing digit, or shift the

flashing digit by one position to the right, by pressing the [PRINT]

key. Confirm your setting by pressing the [TARE] key.

Repeatedly press the [ON/OFF] key until the balance is in weigh-

ing mode.