15.3 Remote control instructions

Attention

:

Inputting characters and command codes not shown here into the balance may not

only alter the previous settings but may also impair proper measurement.

If by mistake characters or commands not shown here are entered into the balance,

immediately unplug the power supply cable and wait about ten seconds before plug-

ging again.

Command Code

Function

Description

D01

Continuous output

The balance continuously outputs every 110ms .

D05

1 time output

Corresponds to [PRINT] key

D06

Auto Print

See 14.4

D07

1 time output with stability

information

The status of the stability mark is appended to the head
of the data with output.
S: when the standstill display appears
U: when standstill display does not appear

D08

1 time output at stability

After command input, the data are output at the first
appearance of the standstill display.

D09

Halt output

Auto Print and continuous output halted

Q

ON/OFF switching

Switches between standby status and measurement
status.

T

Taring

Corresponds to the [O/T] key

TS

Taring after stability wait

After command input, taring is done at the first appear-
ance of the stability mark.

C18

Measurement span calibration

+

Measurements in add-up mode

See ch. 14.5

R

Total reset

All application measurements terminated and reset

mg

mg unit

PERCENT

Percent determination

PCS

Parts counting

CT

ct unit

SDENCE

Density of solids

LDENCE

Density of liquids

%

100% setting

G

g, % switching

- g

g unit removal

- mg

mg unit removal

- PERCENT

Deleting percentage determina-

tion t

- PCS

Piece counting removal

- CT

ct unit removal

- SDENCE

Solid specific gravity removal

- LDENCE

Liquid specific gravity removal