17 instant help – KERN & SOHN ABT 220-5DM User Manual

Page 71

Advertising
background image

ABT-BA-e-1312

71

17 Instant help

In case of an error in the program process, briefly turn off the balance and discon-

nect from power supply. The weighing process must then be restarted from the be-

ginning.
Error code table:

Error code display

Explication

Remedy

CAL E2

High zero point shift dur-

ing adjustment

Remove items from the

weighing pan.

CAL E3

Great deviation of meas-

uring values in PCAL.

Use correct adjusting

weight.

CAL E4

Great deviation of meas-

uring values during ad-

justment

CHE X (X is a numeral)

(when the display stops

here)

Internal malfunction

Please contact a service

representative.

Err 0X (X is a numeral)

Internal malfunction

Please contact a service

representative.

Err 20

An improper value setting

was attempted.

Enter the correct numbers

or decimal points.

Err 24

Power voltage error

Check the power voltage .


Possible causes of errors:

When

Symptom

Possible causes

Remedy

Before

meas-

urement

Nothing appears in the

display.

The AC adapter is not securely

connected. The power switchboard

of the room is turned off. The power

voltage is incorrect.

Check power supply and

connect correctly.

During

meas-

urement

The display fluctuates.

The standstill display

does not appear fast

enough.

The measured results

have poor repeatabil-

ity.

“CAL d” appears fre-

quently.

Vibrations or draft.

Change the installation site.

Change stability and reac-

tion settings or standstill

display.