Procedure

 Confirm menu setting

by

.

Ensure that there are no objects on the weighing plate.

 Wait for stability display, then press

.

 The currently set adjustment weight will be displayed.

 Either use the displayed adjustment weight or change it with

,

and

(numerical input see chapter 2.1.1), the

active digit is flashing.

 Confirm by

, „LoAd“ will be shown.

 Carefully place adjusting weight in the centre of the weighing

plate. Wait for stability display, then press

.

 After the adjustment the balance will carry out a self-test.

Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.