Annex page 2 of 7

Annex to Test Certificate No. DK0199-R76-11.10

Issued by DELTA

Weighing unstable samples

Totalisation

3.

Technical data

3.1

Indicator

Type

KFA-TM / KFE-TM / KFC-TM

Accuracy class

III or IIII

Weighing range

Single-interval, multi-range or multi-interval

scale intervals (n)

6000 or 2

×3000 for class III,

1,000 or 2

×1000 for class IIII

Minimum input voltage per VSI

1 µV

Maximum capacity of interval (Max

i

): n

i

× e

i

Verification scale interval, e

i

=

Max

i

/ n

i

Initial zero-setting range:

± 10 % of Max

Maximum tare effect:

100 % of Max

Fractional factor (pi)

0.5

Excitation voltage

5 VDC

Circuit for remote sense

Not active

Minimum input impedance

350 ohm

Maximum input impedance

1200 ohm

Connecting cable to load cell(s):

See Section 3.1.1

Supply voltage:

9-12 VDC

Operating temperature range

Min / Max = -10 °C / +40 °C

Peripheral interface(s)

See Section 4

3.1.1

Connecting cable between the indicator and the junction box for load cell(s),
if any

3.1.1.1 4-wire system

Line

4 wires, shielded

Maximum length

The certified length of the load cell cable, which shall be con-

nected directly to the indicator.

4.

Interfaces

4.1

Load cell interface

Refer to section 3.1.1.

Any load cell(s) can be used for instruments under this certificate provided the following conditions
are met:

There is a respective test certificate (EN 45501) or an OIML Certificate of Conformity (R60)
issued for the load cell by a Notified Body responsible for type examination under the
Directive 2009/23/EC.

The certificate contains the load cell types and the necessary load cell data required for the
manufacturer’s declaration of compatibility of modules (WELMEC 2, Issue 5, 2009, section