9.4 Error messages

Error

message

Description

Possible causes

- - - - -

Maximum load exceeded

Unload weighing system or reduce

preload.

- - ol - -

Err 1

Incorrect data input

Follow format “yy:mm:dd“

Err 2

Incorrect time entry

Follow format “hh:mm:ss“

Err 4

Zeroing range exceeded due
to switching-on balance or

pressing

(normally 4% max)

Object on the weighing plate

Overload when zeroing

Err 5

Keyboard error

Err 6

Value outside the A/D
changer range

Weighing plate not installed

Damaged weighing cell

Damaged electronics

Err 9

Stability display does not
appear

Check the environmental

conditions.

Err 17

Taring range exceeded

Reduce load

Fai l h /

Fai l l

Adjustment error

Repeat adjustment.

Ba lo /

Lo ba

Battery very low

Recharge battery


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.