Т³гж 17 – LABEC AA-17BS User Manual

Page 16

Advertising
background image

SEA Series Steam Sterilizers

Operation manual

12

7.2 USB module function

Caution:

This function is optional to the demand of customer.

7.2.1Main function:
Storage of printed data of sterilization
Built-in FLASH memorizer
U-disk. Download

7.2.2Basic parameters:
Storage Capacity: 2M bytes
Quantity of printed data could be stored: 2048
Support FAT files

7.2.3Operation instruction:
1.Turn on the power, the mode of USB will start automatically. After self-test, it goes into alert
with the indicator flickering-1 time.
2.After the sterilization cycle ends, the data of sterilizer is sent to the USB mode(The printer
program should be turned on). Indicator will flicker-1 time after the data-transfer finish.
3.During the alert, insert U-disk. The mode indicator will be on, with the U-disk indicator
flickering. And now, it is in the process of data-download. Don't turn the power off or draw
out the USB-disk, which will cause the data lose.
4.After the download finish, USB mode indicator will flicker twice, reminding the user to draw
out the U-disk.
5.During the sterilization cycle, avoid data download, or else, the data will lose.
6.Connect U-disk to computer. You will see files like _2000_, where the downloaded data is kept,
following turns of Year, Month and Day. Easy to print it out.
7.USB mode terror(According to the flickering times of USB mode indicator) and solution:

Times

1

2

3

4

5

6

7

Definition

Solution

Self-test passed

Operation finish

No definition now.

U-disk needs formation

Not enough capacity

U-disk needs formation

USB mode error

Normal

Normal

No definition now.

Format U-disk in FAT files

Delete the files in U-disk

Format U-disk in FAT files

Contact us

Advertising