SEA Series Steam Sterilizers

Operation manual

23

13.Appendix.

Form 1 Work mode chart

Form 2 Error code chart(In case of faults, please contact your distributor or the immediately)

Sterilizing temperature

Sterilizing pressure

Class

Pre-heating

Vaeuuning time

Sterilizing time

Drying time

Quick drying time

Total time

B cycle

134

B&Dtest

Admission time

134

134

121

134

2.10

2.10

2.10

1.10

2.10

B

B

S

B

B

0-7min

0-7min

0-7min

0-7min

0-7min

10min

10min

10min

10min

4min

4min

5min

5min

5min

5min

5min

4min

18min

4min

16min

3.5min Keep pressure

10min

10min

10min

10min

10min

1min

5min

5min

5min

5min

36(31)min 50(45)min

30(25)min

48(43)min

19.5min

15min

1

2

3

4

5

6

7

8

Item

9

Error

sensor error

super pressure >2.6Bar

o

super temperature >145 c

Constant temperature fails

block of steam pipe (20S<0.3Mba)

failure of the door

Steam generator over

temperature (>230 C)

heating overtime (>25 min)

pressurization

overtime (>60 min)

Solution

Examine inner temperature

sensor or pressure sensor

examine outer temperature

sensor

Return parameter autoadjus

programm

Examine the steam output

valve or vacuum valve

examine the door switch

Check steam generator

temperature sensor

examine the electronic connection

of heat ring and heat stick

examine the steam input pipe

examine sensor

18min

Bcycle

Unwrappd

cycle

RUB & PLAS

cycle

Vacuum test

Clean test

Clean 20

min

10

min

30

min

Drying

The figures in the chart are only for reference, the sterilization effect verified by user, and the user can
choose the sterilization time according to the verification results. Which cycle should be adopted is
according to the type of the instruments.