15.Important information

1.The autoclave must be installed on a leveled surface.

2.When operated at an elevation of more than 500 meters the machine must be
reprogrammed. Please get in touch with your distributor for instructions.

3.Distilled water must be used for prolonging the life of the machine.

4.The ventilation openings in the machine must not be blocked or covered.

5.Instruments should be placed on tray with space between one another
allowing steam to pass through.

6.Condensed water collector must be emptied from time to time.It should be emptied
when refilling the water reservoir.(Draining of the collector can be followed as
per item 9.2)

7

.Pls closing the door well while you use it.

8.Never try to open the chamber door if and when the pressure display doesn't show 0.0 .

9.Before any kind of maintenance turn off the power to the autoclave and allow it to cool down.

10.DO NOT drag the autoclave when it is being moved.

11.Power connection must be grounded properly.

12.Must be provided enough power (22L:2300W; 17L:1800W).

13.If the ambient temperature is low please pre-heat the autoclave without instruments for 30
minutes before sterilizing.

14.DO NOT power off the unit during sterilization,.
To abort a cycle during sterilization press the Start/Stop key for 3 seconds.

15. The chart in the operation manual only for reference, which keep the actual product for real.