SEA Series Steam Sterilizers

Operation manual

Fig 3-2

3.Installation

The Autoclave should be installed in a ventilated place allowing minimum of 10 cm space at all
sides and 50 cm at the top. The ventilation openings at the side of the machine should not be
blocked. The Autoclave should be placed on a leveled counter top as Fig 3-2. The dimension of
the machine are shown in Fig 3-1:

When moving the unit DO NOT DRAG. Lift the antoclave to move.

If the front of the machine is lower than the back the control will display

E3

(E3 means it

will not dry completely). Adjust the height by adjusting foot pad of the front of Autoclave.

Caution

3

585

585

**The front of the autoclave must be higher than the rear.
**Side to side the autoclave must be level.
**Use level supplied to achieve correct alignment see (FIG 3-2)

Fig 3-1