Т³гж 32, Trouble shooting – LABEC AA-17BS User Manual

Page 31

Advertising
background image

SEA Series Steam Sterilizers

Operation manual

27

14.Trouble shooting

1.During venting or drying if a thump sounds it means that the Condensed water collector is full.
You must drain the water from the tank(see section 9.2 , Fig 9-4).

2.During a cycle if a banging sounds, steam leaks from the chamber door and the displays show
an

E6

code. Please verify whether the door knob is properly turned and locked. If the

steam leaks at the beginning of a cycle, you will need to change the seal according to
instructions at section 9.6.

3. At the end of a cycle if, there are water drops on the instruments.
You can correct the problems by:

1).Checking the drain filter is not blocked (see section 9.2).
2). Reduceing the load of the instrument in the chamber.( especially the porous and dressing
and other water absorbent materials.)
3). Waiting for a while at the end of a cycle

Ed

for better drying result. The autoclave

will continue heating the chamber until you open the door.

Advertising
This manual is related to the following products: