LABEC M-MIA-CC User Manual

Page 40

Advertising
background image

Ultrasonic milk analyser

Operation manual

23.08.10

38/98

Recalibrated

Analyser ready


This means that the calibration was completed successfully and the analyser
is recalibrated for cow milk, marked as “Calibr 1”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.

If during work the sample’s temperature exceeds the temperature range 15-
25° С the following message appears:

Temperature out of

range


Wait till the end of the measurement. When the message appears:

Put sample again


Put sample with temperature in the temperature range and continue the
measurement till completion of 5-time measurement.