LABEC M-MIA-CC User Manual

Page 46

Advertising
background image

Ultrasonic milk analyser

Operation manual

23.08.10

44/98

If after recalibration “Save calibration” is pressed the new calibration
settings will be saved over the factory settings. After that is impossible
to restore the factory settings of the calibration. Save the newly made
calibration only if you are sure about its correctness.

7.2.3.7. Settings Page 2.

After this menu is started the display shows the following:


Settings Page2
Final clean Cnt
Auto Print Res
Larg Res En/Dis
Set Calibr Name

Exit


Now there is a possibility one of the following options to be set:

7.2.3.7.1. Final Clean Cnt
Sets the number of cleaning cycles.

7.2.3.7.2. Auto Print Res.
It is possible (if it is needed) to prohibit the automatic printing the measurement
results. If there is external keypad connected, then the results could be printed
by pressing the button 1 from the keypad. It is possible to print out unlimited
number of printouts for one and the same measurement.

7.2.3.7.3. Larg Res En/Dis.
The format of the measurement data send towards the computer is set.
If the option Large Disable, is chosen, then only the main results are send to
the computer – Fat, SNF, Density, Lac, Proteins, Added Water, sample
temperature, device serial number and calibration number.
If the option Enable, is chosen, except the above mentioned parameters also
data for Salts, Freezing Point, pH, Conductivity will be send to the computer. In
this case is necessary the software in the computer to be conformable to the
format of the sent data. After starting the menu, the display shows (for example):