LABEC M-MIA-CC User Manual

Page 49

Advertising
background image

Ultrasonic milk analyser

Operation manual

23.08.10

47/98

There are the following possibilities:
With button Exit – to leave the menu.
With button Save – to confirm already edited calibration name and to save it in
the device. The program returns to the beginning of the menu for setting
calibration names.


7.2.4. Tests.

Start different tests. Possibilities:

7.2.4.1. Test pump.
Starts pump’s test. The number of the completed suction/display cycles is
indicated.

7.2.4.2. Ultrasound.
Test for the ultrasonic system. Used in production conditions.

7.2.4.3. Serial Prn.
Display a short text of a serial printer, connected to COM2 – output with
inscription Printer on the back panel of the device.

7.2.4.4. Set Amplitude.
Serves for ultrasound amplitude adjustment. It is used under production
conditions or by the customer (after sensor change) according the
instructions in the document SetCell.pdf.

Please, use this menu only after reading the above pointed document
SetCell.pdf

7.2.5. pH meter & Co meter

Ph and conductivity measuring are additional possibilities for the analyser
and are optional. Their usage is described in Appendices PH Measurement
and Conductivity Measurement.

7.2.6. Exit

By pressing the button you may leave the program and pass towards another
menu.

Advertising