During operation

Maximum ambient temperature: 40 °C

• Make sure ventilation is adequate if pump is installed in a housing or if ambient

temperature is elevated.

Potentially dangerous gases or vapours at the outlet of the pump have to be

drained and disposed appropriately.

• Due to the high compression ratio of the pumps, pressure at the outlet port might

be generated being higher than the maximum permissible pressure compatible

with the mechanical stability of the system. Ensure that the pump outlet cannot

become blocked or restricted.

If pump is installed in altitudes of more than 1000 m above mean sea level check

compatibility with applicable safety requirements, e. g. IEC 60034 (motor may over-

heat due to insufficient cooling).

Do not start the pump if the

pressure at outlet port exceeds maximum 1.1 bar

(absolute). Attempts to start the pump at higher pressures may cause blockade and

damage of the motor.

Check compatibility with

maximally permitted pressure at inlet and outlet.

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves

will be damaged, if liquids are pumped in significant amounts.

Check the pump regularly for external soiling and deposits, clean if necessary to

avoid an increase of the pump‘s operating temperature.

If pumping condensable vapours (water vapour, solvents, ....), let the pump run with

gas ballast to reduce condensation in the pump.

The motor is protected by a

temperature sensor at the circuit board: Current

limitation if the temperature at the circuit board is higher than 70°C, switching off the

pump if the temperature is higher than 85°C. In case of blockade of the motor (after

10 attempts to start-up) the pump is switched off.

If the pump is switched off due to safety measures, manual reset is necessary. Iso-

late the pump from mains. Eliminate cause of failure before restarting the pump.

Avoid high heat supply (e. g. due to hot process gases).

A warm up period (approx. 15 min.) is required to ensure that rated ultimate vacuum

and pumping speed are attained.

Attention: Important notes regarding the use of gas ballast

When using air rather than inert gas, risk of significant damage to equipment and/

or facilities, risk of personal injury or even loss of life exists due to the formation

of hazardous and/or explosive mixtures if air and pumped media react inside or

at the outlet of the pump.

Make sure that air/gas inlet through the gas ballast valve never leads to hazard-

ous, explosive or otherwise dangerous mixtures. If in doubt, use inert gas.