Shutdown, Notice – VACUUBRAND PC 3001 basic User Manual

Page 14

Advertising
background image

page 14 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999198 / 29/09/2009

In case of low boiling solvents when the formation of condensate is unlikely, the use

of gas ballast might be unnecessary.

Shutdown

Short-term:

Has the pump been exposed to

condensate?

- Allow the pump to continue to run at atmospheric pressure for a few minutes.

Has the pump been exposed to media which may damage the pump materials or

forms

deposits?

- Check and clean pump heads if necessary.

Long-term:

- Take measures as described in section short-term shutdown.

- Separate pump from the application.

- Close gas ballast.

- Drain catchpots (if using with upgrade kits)

- Close inlet and outlet port (e. g. with transport caps).

- Store the pump in dry conditions.

In case of

condensable vapours (water vapour, solvents,

...):

Do not pump vapour until the pump has reached its opera-

ting temperature.

Open gas ballast valve. The gas ballast valve is open if the

arrow on the gas ballast cap shows towards the pump.

With gas ballast valve open ultimate vacuum will be re-

duced, pumping speed is decreased.

Use inert gas at the air inlet to avoid the formation of explo-

sive mixtures.

Close gas ballast valve by turning 180°.

gas ballast

NOTICE