Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999198 / 29/09/2009

Section ” Replacing the diaphragm” see above.

Disassembling the housing cover at the side of the on/off switch

Lay the pumping unit down with the housing cover at the

side of the ON/OFF switch facing upwards (support pump-

ing unit appropriately).

Use a hex key size 4 to loosen the six socket head screws

at the pump head.

Raise the housing cover.

Pull off the hose from the housing cover. Do not kink the

hose.

Remove head covers from housing cover and check

valves. Note position of valves and remove.

Never remove parts by using a spiky or sharp-edged tool

(e. g. screw driver), we recommend to use a rubber mallet

or compressed air (to be blown carefully into port).

Replace valves if necessary.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.