page 21 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999198 / 29/09/2009

scheme pump head with head covers and valves

valves at the outlet

(round centred

opening under

valve)

valves at the inlet

(kidney-shaped

opening beside

valve)

”side with hose connection” between pump heads

”side with knob” / stand foot

Assembling the housing cover at the inlet side

Assemble head covers and valves.

Check for correct position (see also fig. below).

It is absolutely essential to comply with the positions and orientations of the

head covers and valves!

Position housing cover.

Move housing cover slightly to make sure that the head

covers are correctly positioned.

Screw in the six socket head screws fixing the housing

cover crosswise first slightly, then tighten.

Do not tighten until head cover is in contact with housing,

max. torque 6 Nm.