Use an open-ended wrench (w/f 17) to loosen the union

nut at the hose connection next to the gas ballast.

Use an open ended wrench (w/f 14) to turn the elbow fit-

ting 1/4 of a turn, disconnect the hose. Do not remove the

elbow fitting from the pump head.

Through reassembly a leak may result.

Use hex key size 5 to unscrew the handle from one side of

the housing cover.

Disassembling the housing cover at the inlet side

Lay the pumping unit down with the housing cover at the

side of the pump inlet facing upwards (support pumping

unit appropriately).

Use a hex key size 4 to loosen the six socket head screws

at the pump head and remove the upper housing (housing

cover with housing cover insert). Do not kink hoses.

Never remove parts using a spiky or sharp-edged tool (e.

g. screw driver), we recommend to use a rubber mallet or

compressed air (to be blown carefully into port).

Take the head covers carefully off the housing cover to

check the valves.

Note the position of the valves and remove them.

Replace valves if damaged.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.