Technical data

*

Information: In a motor speed range between

0 and 200 rpm the pump is running automatically in a

cycled interval operation.

Type

PC 3001 basic

Maximum pumping speed (ISO 21360)

m

3

/h

1.7

Ultimate vacuum (absolute) without gas ballast

mbar

2

Ultimate vacuum (absolute) with gas ballast

mbar

4

Maximum permissible inlet pressure (absolute)

bar

1.1

Maximum permissible outlet pressure (abso-

lute)

bar

1.1

Maximum permissible pressure (absolute) at

gas ballast valve

bar

1.2

Permissible ambient temperature storage / op-

eration

°C

-10 to +60 / +10 to +40

Permissible relative atmospheric moisture dur-

ing operation (no condensation)

%

30 to 85

Rated power (electric)

kW

0.16

No-load speed*

min

-1

0 - 2400

Maximum permissible range of supply voltage

(

±10% ) Attention: Observe specifications

of rating plate!

100-120 V~ 50-60 Hz

200-230 V~ 50-60 Hz

Maximum rated current at:

100-120 V~ 50/60 Hz

200-230 V~ 50/60 Hz

A

A

1.6

0.7

Device fuse

1x 7A

1x 0.5A self-resetting (VACUU•BUS)

Motor protection

temperature sensor on the pcb

(current limitation)

Degree of protection IEC 529

IP 20

Inlet

hose nozzle DN 6/10 mm

Outlet

hose nozzle DN 10 mm / silencer

Dimensions L x W x H approx.

mm

251 x 256 x 400

Weight approx.

kg

6.4

Advertising