Maintenance and repair, Notice – VACUUBRAND PC 3001 basic pro User Manual

Page 19

Advertising
background image

page 19 of 54

A temperature sensor at the circuit board protects the mo-

tor: Current limitation in case the temperature at the circuit

board raises above 158°F (70°C). At temperatures above

185°F (85°C) the pump is switched off. In case of a motor

blockage (after 10 start-up attempts) the pump is switched

off.

Note: Only manual reset is possible. Disconnect the pump

from the power source. Identify and eliminate the cause of

failure.

Maintenance and repair

In order to comply with laws (occupational, health and

safety regulations, safety at work law and regulations for

environmental protection) vacuum pumps, components

and measuring instruments can only be returned when

certain procedures (see section „Notes on return to the

factory“, pg. 48) are followed.

Take advantage of our service seminars, which put special

focus on the maintenance and repair of vacuum pumps.

For details and for the online ”Instructions for repair” man-

ual see www.vacuubrand.com.

In normal use, the lifetime of the diaphragms and valves is

typically 15,000 operating hours. Bearings have a typical

durability of 40000 h.

Ensure that the pump cannot be operated acciden-

tally. Never operate the pump if covers or other

parts of the pump are disassembled.


Switch off the pump. Disconnect the electrical

power cord and wait two minutes before starting

maintenance to allow the capacitors to discharge.

Note: The pump may be contaminated with process

chemicals, which have been pumped during operation.

NOTICE

Advertising