Make sure that air/gas intake through the gas ballast

valve can never lead to hazardous, explosive or other-

wise dangerous mixtures. If in doubt, use inert gas.

To reduce condensation in the pump, do not pump vapor

before the pump has reached its operating temperature.

Open the gas ballast valve when pumping condensable

vapors. Turn gas ballast cap to open valve.

For

condensable vapors (water vapor, sol-

vents, etc.):

- The gas ballast valve is open if the arrow on

the gas ballast cap is pointing towards the

pump.

- With gas ballast valve open, the ultimate

vacuum will be reduced.

- Use inert gas for gas ballast to avoid the formation of

explosive mixtures.

Attention: maximum supply pres-

sure of inert gas: 17.5 psi (1.2 bar) absolute.

- Close the gas ballast valve by turning the cap 180°.

In case of low boiling solvents (when the formation of con-

densate is unlikely), the use of gas ballast might be un-

necessary.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits?

- Check and clean pump heads if necessary.