Assembling the fittings, Checking the ultimate vacuum – VACUUBRAND PC 3001 basic pro User Manual

Page 46

Advertising
background image

page 46 of 54

Slip connection hose onto hose connector

below the housing cover.

Close hose clip with flat pliers.

Assembling the fittings

Attach the handle at the housing cover (use

5 mm wide Allen key)

Slip the hose onto the hose connector by

turning the fitting (L) with an open-ended

wrench w/f 14.

Tighten the union nut (K) of the hose con-

nection with an open-ended wrench w/f 17.

+

Tighten the union nut first by hand and then

tighten one full turn using the open ended

wrench.

Checking the ultimate vacuum

After any intervention at the equipment (e.g., repair / maintenance)

the ultimate vacuum of the pump has to be checked. Only if the pump

achieves its specified ultimate vacuum, the pump’s leak rate is low

enough to ensure that no explosive atmospheres will occur in the inte-

rior of the equipment.