page 24 of 54
Type
PC 3001 basic
Inlet
hose nozzle for tubing I.D.
1/4” / 3/8”
(hose nozzle DN 6/10 mm)
Outlet
hose nozzle for tubing I.D. 3/8”
(hose nozzle DN 10 mm) /
silencer
A-weighted emission sound pressure
level** (uncertainty K
pA
: 3 dB(A))
dB(A)
42
Dimensions L x W x H
approx.
in
(mm)
9.9 x 10.1 x 15.7
(251 x 256 x 400)
Weight approx.
lbs.
(kg)
14.1
(6.4)
We reserve the right for technical modification without prior notice!
*
Running smoothly only at motor speeds higher than 200 rpm
**
Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 1500rpm and
ultimate vacuum with exhaust tube at outlet.
Gas inlet temperatures
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F*
(-10°C to +80°C*)