Make sure ventilation is adequate, especially if the pump

is installed in an enclosure, or if the ambient tempera-

ture is elevated. Provide external ventilation, if neces-

sary.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Note: Flexible elements will compress or flatten when

evacuated if not designed for use under vacuum.

• Hose connections at the pump inlet must always be

gas tight.

• A power failure may cause accidental ventilation of the

pump, especially if the gas ballast valve is open. If this

constitutes a potential source of danger, take appropri-

ate safety measures.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

The ON/OFF switch is at the left side of the pumping unit.

After switching off the pump, wait minimum 60 sec. before

switching on again.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system.

Use connecting hoses with large diameter and keep them

as short as possible to avoid flow losses. Locate the pump

as closely as possible to the application.

Always install outlet tubing descending from the pump or

provide other measures to avoid backflow of condensate

towards the pump.