Use and operation, Installing a pump in a vacuum system – VACUUBRAND PC 3001 basic pro User Manual

Page 28

Advertising
background image

page 28 of 54

Use and operation

Notes regarding the selection knob and the motor speed

Turning the knob to the left decreases the motor speed, turning the knob

to the right increases the motor speed.

Pumping at

high motor speed increases the pumping speed of the

pump. Ensure sufficient cooling of the pump!

Pumping at

low motor speed increases the lifetime of diaphragms and

valves!

The pump attains the best ultimate vacuum in the low speed range be-

tween approx. 600 and 1000 rpm.

Installing a pump in a vacuum system

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump outlet if

necessary. Always vent exhaust gases appropriately

(e.g., into a fume hood).

+

Never block the gas outlet (hose nozzle 3/8”

(

10 mm) /

silencer). The exhaust line must always be free of ob-

structions (no back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the outlet tub-

ing must be at least the size of the pump’s exhaust

connection.

+

Operation with silencer at the outlet: Check the silenc-

er regularly for chemical attack and deposits, replace

if necessary or use an exhaust line without silencer.

+

Particles and dust must not be aspirated. If necessary,

you must install appropriate filters. You must ensure

their suitability concerning gas flow, chemical resis-

tance and resistance to clogging prior to use.