Assembling the housing cover at the inlet side, Disassembling the housing cover, At the side of the on/off switch – VACUUBRAND PC 3001 basic pro User Manual

Page 44

Advertising
background image

page 44 of 54

Put on housing cover (A).

+

Move housing cover slightly to ensure that the head covers are cor-

rectly positioned.

Screw in the Allen head screws at the head

cover in a diagonal pattern with a 4 mm

wide Allen key, loosely at first, to align, then

tighten.

+

Maximum torque: 4.4 ft

.

lb

f

(6 Nm).

Assembling the housing cover at the inlet side

Disassembling the housing cover at the side of the ON/

OFF switch

Opening the hose clip:

Apply slotted screwdriver as shown and

turn.

Detach the coupling of the hose connection

(M) below the housing cover (A).

Open the hose clip (N) with a slotted screw-

driver.

N

M

A

Lay the pumping unit on its side with the

pump heads to be maintained at the top.

Support the pumping unit appropriately.

Disassemble the housing cover (A) to check

the valves (B).

Unscrew six Allen screws with a 4mm wide

Allen key. Remove the housing cover with

housing cover insert (A) together with head

covers (C) and valves (B).

Advertising