Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014

page 11 of 40

Consider interactions and chemical reactions of the pumped media. Ensure

that the materials of the pump’s wetted parts are compatible with the pumped

substances, see section „Technical data“.

When changing the substances pumped, we recommend purging the pump with

air or inert gas prior to changing the pumped media. Purging the pump will pump

out residues and it will reduce the possibility of reactions of the pumped substan-

ces with each other and with the pump’s materials.

Safety during operation

Adopt suitable measures to prevent the release of dangerous, toxic, explosive,

corrosive, noxious or polluting fluids, vapours and gases. To prevent any emis-

sion of such substances from the pump outlet, install an appropriate collecting

and disposal system and take protective action for pump and environment.

You must take suitable precautions to prevent any formation of explosive mix-

tures in the pump chamber or at the outlet of the pump. In case, e.g., of a dia-

phragm failure, mechanically generated sparks, hot surfaces or static electricity

may ignite these mixtures. Use inert gas for gas ballast or venting, if necessary.

Drain appropriately or otherwise remove any potentially explosive mixtures at the

outlet of the pump, or dilute them to non-explosive concentrations.

Prevent any part of the human body from coming into contact with vacuum.

Make sure that the exhaust pipeline cannot become blocked.

Comply with applicable regulations when disposing of chemicals. Take into con-

sideration that chemicals may be contaminated. Take adequate precautions to

protect people from the effects of dangerous substances (chemicals, thermal de-

composition products of fluoroelastomers). Use appropriate protective clothing

and safety goggles.

Interruption of the pump (e.g., due to power failure), failure of connected compo-

nents or of parts of the supply, or change in parameters must not be allowed to

lead to dangerous conditions. In case of a diaphragm failure or in case of a leak

in the manifold, pumped substances might be released into the environment or

into the pump housing or motor.

Comply with all notes regarding proper use of the pumps, as well as operation

and maintenance guidance.

The residual

leak rate of the equipment might render possible an exchange of

gas, albeit extremely slight, between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of the pumped substances or

the environment.

• Pumping at high inlet pressure may lead to overpressure at the gas ballast valve.

Pumped gases or condensate might be expelled if the valve is open. If an inert

gas supply is connected to the gas ballast, ensure that its inlet pipeline is not

contaminated.

• Pay attention to the safety symbol ”hot surfaces” on the equipment. Hot parts

may cause burns if touched. Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that hot surfaces of the pump do

not cause burns. Provide a suitable contact guard if necessary.

Prevent the backpressure of gases and the backflow of condensates at the outlet.