Bring diaphragm into a position in which diaphragm is in

contact with housing and centred with respect to bore.

Lay the pumping unit down and support it appropriately.

Assemble head cover and valves.

Check for correct position (see fig. below).

It is essential to mind the position and orientation of the head cov-

ers and the valves!

Scheme of pump head with head covers and valves

valves at the outlet

(round centred opening

under valve)

valves at the

inlet

(kidney-shaped

opening beside

valve)

”side with hose connection”

”motor side”