Technical data

Type

PC 301 / MD 1C + 2AK

Maximum pumping speed

at 50/60 Hz (ISO 21360)

m

3

/h

1.3 / 1.5

Ultimate vacuum (absolute) without gas ballast

mbar

2

Ultimate vacuum (absolute) with gas ballast

mbar

4

Maximum permissible inlet pressure (absolute)

bar

1.1

Maximum permissible outlet pressure (abso-

lute)

bar

1.1

Maximum pressure difference between inlet

and outlet

bar

1.1

Maximum permissible pressure (absolute) at

gas ballast valve

bar

1.2

Permissible ambient temperature

storage / operation

°C

-10 to +60 / +10 to +40

Permissible relative atmospheric moisture

during operation (no condensation)

%

30 to 85

Maximum permissible installation altitude

above mean sea level

m

2000

Rated power

kW

0.08

No-load speed at 50/60 Hz

min

-1

1500 / 1800

Maximum permissible range of supply voltage

( ±10% )

Attention: Observe specifications of rating

plate!

200-230 V~ 50/60 Hz

120 V~ 60 Hz

Maximum rated current at:

120 V~ 60 Hz

200-230 V~ 50/60 Hz

A

A

1.7

0.8 / 0.85

Motor protection

thermal cutout, manual reset;

MD 1C C/US: additionally fuse 2.5 AT

Degree of protection IEC 529

IP 40

Inlet

hose nozzle DN 6/10 mm

Outlet

hose nozzle DN 6/10 mm

A-weighted emission sound pressure level*

(uncertainty K

pA

: 3dB(A))

db(A)

45

Dimensions L x W x H approx.

mm

316 x 239 x 460

Weight approx.

kg

10.2

We reserve the right for technical modification without prior notice!

*

Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and at ultimate vacuum with exhaust

tube at outlet.