Check the liquid level in the catchpots regularly and drain condensate in time to

prevent overfilling.

Provide appropriate protective measures to allow for the possibility of failure and

malfunction. The protective measures must also allow for the requirements of the

respective application.

In case of overload, the motor is shut down by a

self-hold thermal cutout in the

winding.

Note: Only manual reset is possible. Switch off the pump and disconnect from the

power source. Identify and eliminate the cause of failure. Wait approx. five minutes

before restarting the pump.

Note: In case of supply voltage below 100V, the lock of the cutout might be

impaired and the pump may restart on its own after sufficient cooling down. Take

appropriate precautions, if an automatic restart of the pump may lead to a danger-

ous situation (e.g., switch off the pump and disconnect from the power source).

Maintenance and repair

In normal use, the lifetime of the diaphragms and valves is typically 15,000 operating

hours. Bearings have a typical durability of 40000 h. Motor capacitors have a typical

durability in the range of 10000 to 40000 h depending strongly on operation condi-

tions including ambient temperature, humidity or load.

Ensure that the pump cannot be operated accidentally. Never operate the

pump if covers or other parts of the pump are disassembled.

Switch off the pump. Disconnect the electrical power cord and wait five

seconds before starting maintenance to allow the capacitors to discharge.

Note: The pump may be contaminated with process chemicals, which have been

pumped during operation. Ensure that the pump is completely decontaminated

before maintenance commences.

Take adequate precautions to protect people from the effects of dangerous sub-

stances if contamination has occurred. Use appropriate protective clothing, safe-

ty goggles and protective gloves.

Wear parts have to be replaced regularly.

Never operate a defective or damaged pump.

Check every motor capacitor regularly by measuring its capacity and estimating

its time in operation. Replace old capacitors early enough to prevent a failure in

operation. If an old motor capacitor fails, the capacitor may get hot. It may even

melt or emit a flame, which could be

dangerous for persons and equipment in

the vicinity. The capacitors have to be replaced by an electrician.

Vent the pump before starting maintenance. Isolate the pump and other com-

ponents from the vacuum system. Allow sufficient cooling of the pump. Drain

condensate, if applicable.

Ensure that

maintenance is done only by suitably trained and supervised techni-

cians. Ensure that the maintenance technician is familiar with the safety procedures

which relate to the products processed by the pumping system.

In order to comply with laws (occupational, health and safety regulations, safety at

work law and regulations for environmental protection) vacuum pumps, components

and measuring instruments can only be returned when certain procedures (see sec-

tion „

Notes on return to the factory“) are followed..