Cleaning and assembling components – VACUUBRAND PC 301 User Manual

Page 35

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014

page 35 of 40

Round bottom flask 500 ml, coated .........................638497

O-ring 28 x 2.5 ...........................................................635628

at the spherical ground joint of the round bottom flask

Cleaning and assembling components

Flow control diaphragm ...........................................677137

Cleaning and inspecting the manual flow control dia-

phragm:

Unscrew union nut.

Remove existing deposits, if necessary with cleansing

agent.

Assembly of flow control diaphragm

Fix diaphragm by hand and turn control button in direction

of arrow + (open). Open flow control diaphragm complete-

ly.

Do not use any tools to fix the diaphragm to avoid damag-

ing the diaphragm.

Assemble flow control diaphragm with the four marks of

the operating part fitting in the four dents of the valve bloc.

Otherwise leakage will result. Avoid any canting of the

components.

Tighten union nut.

If the pump does not achieve the ultimate vacuum:

- Whenever the diaphragms and valves have been replaced, a break-in period of several hours is

required before the pump achieves its ultimate vacuum.

- In case of an unusual noise, switch off pump immediately and check clamping disc positions.

If the specified ultimate vacuum is not achieved, and if this does not change after the break-in period:

Check hose connectors at pump heads for leaks. If necessary recheck valve seats and pump chambers.