Program function

Permits ten programs to be defined and stored, each with up to

ten program steps with preset values for vacuum and time.

+

Edit:

Use to define the preset values for the process run:

Time: Defines either the process runtime for each program step to

reach a preset vacuum level or, if programming a ”Step”, the runtime

after having achieved the vacuum level. The total process runtime is

shown in the base line. Attention: A preset runtime of 99:59:59 hours

in the final program step will cause the process to run endlessly. Ter-

minate the process by pressing the ”STOP” key .

Vacuum: Vacuum level to be attained.

Venting valve: Operating a venting valve to reach a preset vacuum

level.

”Step”: ”Step” causes pump down as fast as possible to the preset

vacuum level. As soon as the vacuum level is reached the time meter

starts running.

”Auto” (only with NT VARIO / VARIO-B pumps): Auto =

: indicates

that the system will search for a boiling point automatically, and will

automatically adapt to changes in the boiling point until a preselected

time interval expires.

Auto = : indicates that the system will automatically adapt to chang-

es in the boiling point. The program step is finished when the selected

time or pressure value has been reached.

Selecting ”Auto” in combination with ”Step” in one program step is not

possible.

+

Open: Choose program (Programs 0 - 9).

+

Store: This command stores an edited program or the program of the

last process to one of the storage spaces 0 - 9.

+

Hysteresis: The ”Hysteresis” is the control bandwidth of the two-point

control. A too small hysteresis will lead to frequent switching of the

valve or the pump. A too large hysteresis will lead to imprecise con-

trol. Suggested ”Hysteresis” values are stored in the controller (setup