alternatively:

For diaphragm pump with in-line valve and/or Vacuum-Management-

System

+

Select ”

Vac control” function.

+

Set ”Set vacuum” (and ”Hysteresis”, if necessary) depending on the

solvent and the bath temperature.

+

Considering the hysteresis and the bath temperature, set ”Set vacu-

um” to a value which ensures that the solvent will definitely boil.

+

To set ”Maximum” is usually not necessary, unlike filtrations, because

the pressure does not increase at the end of the evaporation.

+

Set a value for ”Duration” if the process should be terminated auto-

matically after a definite time.

+

Use ”Delay” to pump out condensate and clean the pump at the end of

the process. The in-line valve is closed and so the pump is separated

from the application. During ”Delay” the coolant valve is still open.

+

Start process by pressing ”START/STOP” key.

alternatively:

For NT VARIO / VARIO-B diaphragm pumps (vacuum control with

pinpoint precision)

+

Select ”

Vac control” function.

+

Set a maximum ”Speed”, to select a faster or slower process.

+

To set ”Maximum” is usually not necessary, unlike filtrations, because

the pressure does not increase at the end of the evaporation.

+

Set a value for ”Duration” if the process should be terminated auto-

matically after a definite time.

+

Set ”Set vacuum” depending on the solvent and the temperature.

+

Start process by pressing ”START” key.

+

To temporarily switch to ”Auto mode”: Press ”MODE” key. The

controller switches to ”Auto mode”. The setting of the controller does

not change due to this switching. After pressing ”START/STOP” key

the controller is in ”Vac control” again.