This screen-shot shows the factory-set values.

VACUULAN

40 mbar

00:13:00

VACUULAN
Set vacuum 19 Torr

Switch on 150 Torr

Delay

15

min

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

760 Torr

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.
+

Press the selection knob twice to return to the standard display.

+

Switch on (the higher switching value): If the pressure exceeds this

pressure, pumping down starts again.

The ”Switch on” pressure is adjustable in the range of 2-795 Torr (2-

1060 mbar) (at the least 1 Torr (mbar) higher than the ”Set vacuum”).

In the event of a sudden high-pressure spike, pumping starts again

even if the ”Switch on” pressure has not been reached (pressure in-

crease control).

+

Delay: If the vacuum is below ”Set vacuum” for longer than the ”Delay

time, the pump is stopped. Pumping starts again in the event of a rapid

pressure increase or if the ”Switch on” level is exceeded.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the pump stops immediately when the pressure

drops below ”Set vacuum”.).