Adjustment under vacuum – VACUUBRAND CVC 3000 User Manual

Page 53

Advertising
background image

page 53 of 80

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible if the pressure is higher than 525 Torr

(700 mbar).

Vent the measurement connection of the CVC 3000 or in case that of

a connected optional external gauge head VSK 3000. Make sure that

the pressure transducer is at atmospheric pressure.

In ”Configuration” menu, select program ”Adjustment” at the controller.

Use the selection knob to adjust the reading to the current atmospheric

pressure.

Press the selection knob to confirm.

760 Torr

Note: To determine the actual atmospheric pressure, use an accurate

barometer or get accurate reading from the weather service, or a nearby

airport or other reliable source (taking into account the difference in alti-

tude between the source and the laboratory).

Adjustment under vacuum

An adjustment under vacuum is only possible

if the pressure is lower than 15 Torr (20 mbar)

absolute.

Evacuate the measurement connection of the CVC 3000 or in case

that of a connected optional external gauge head VSK 3000 to a pres-

sure < 0.1 Torr (mbar) (e.g., by applying a good two-stage rotary vane

pump).

In ”Configuration” menu, select program ”Adjustment” at the controller.

+

The reading is automatically adjusted to ”zero”.

Press the selection knob to confirm.

0 Torr

Note: Adjustment under vacuum with an actual pressure higher than 0.1

Torr (mbar) reduces the accuracy of the measurement. If the pressure is

significantly higher than 0.1 Torr (mbar), adjustment to a reference pres-

sure is recommended.