The screen-shot shows the factory-set values.

Editing:
+

To select row: turn and press selection knob.

+

To adjust parameter: turn the selection knob.

+

To confirm parameter: Press selection knob. Controller will accept

change and jump to the next parameter in the same row.

+

After 5 seconds without a change, the parameter is assumed to be the

current setting. Select the next row to edit or return to the Program menu.

In case, store an edited program after having quit ”Edit” by storing it to

one of the storage spaces 0 - 9 (select ”Store”).

1013 mbar

00:28:00

Program 5 : 4 1013mbar

Program
Edit

Open

0

Store

0

Hysteresis

Auto

Delay

Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

760 Torr

When selecting ”Graphic”, the display shows a pressure vs. time curve.
Program number, and the step number in that program, along with the

vacuum setting, the actual current pressure and the actual runtime are

displayed across the top.

”Auto”) and are adapted automatically to the preset pressure. The hys-

teresis can be adapted at any time.

The ”Hysteresis” is adjustable in a range of 1-225 Torr (1-300 mbar) or

can be set to ”Auto”.

NT VARIO / VARIO-B pumps are controlled without hysteresis.

+

Delay: Delay” determines the time the coolant valve remains open af-

ter the process has been stopped. Determines also the time the pump

(only with VMS module and in-line valve) remains running after the

process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately - and that

a pump with VMS module and in-line valve is switched off immediately

- when the process stops.).