4.2 Display at start-up

When power is supplied to the

MX32 C

ONTROL

,

the unit goes through a self-test sequence as follows:

1.

Current software version and buzzer test.

Examples of Display:

120

ver

1 0

m32

2.

Current menu access code.

Example of Display:

1270

See section 5.0 for Maintenance Menus

Section 5.5 for Access Code

3.

Current AL1 threshold displayed for channel 1, while the AL1 and fault light for channel 1 light up.

Example of Display:

20

co

1

4.

Current AL2 threshold displayed for channel 1, while the AL2 and fault light for channel 1 light up.

Example of Display:

40

co

2

5.

Current AL1 threshold displayed for channel 2, while the AL1 and fault light for channel 2 light up.

Example of Display:

19.5

o2

1

6.

Current AL2 threshold displayed for channel 2, while the AL2 and fault light for channel 2 light up.

Example of Display:

23.5

o2

2

7.

Count down (in seconds), time for the sensor(s) to stabilize, and both fault LED are lit.

Example of Display:

0.40

min

This time is programmed at the factory
for best results. See section 5.2 to change
programmed setting.

8.

Then the

MX32 C

ONTROL

will begin the normal display cycle

Examples of Display:

0

co

20.9

o2

See section 4.4 for
Normal Display Cycle