0 maintenance, Enrg, Sa e – ENMET MX-32 User Manual

Page 16

Advertising
background image

ENMET Corporation

MX32

E

NGUARD

12

5.0 Maintenance

5.0 Maintenance

5.0 Maintenance

5.0 Maintenance

The

MX32 C

ONTROL

unit has 5 maintenance menus that are accessed by pressing the MENU switch and entering a

valid access code. The access code is set at the factory and may be changed by following the access code menu
explained in section 5.5.

C

AUTION

:

The procedures and adjustments described in the following sections must be performed by authorized

personal. Failure to follow instructions may jeopardize accurate measurements.

See section 4.1 for a description of the pushbutton switches

NOTE:

There may be differences in the display for request to save what has been entered in the maintenance menus.

Example:
May display
ENRG or SAVE

enrg

oui

Example: May display:

OUI or YES
NON or NO

sa e

yes

5.1 Enter the maintenance menus

When the MENU/ENTER switch is pressed the

MX32 C

ONTROL

unit will request an access code to enter the

maintenance portion for programming, initialization, calibration, changing access code and buzzer setting.

If no switch is pressed for one minute, the

MX32 C

ONTROL

will return to normal display mode.

5.1.1 Entering the Maintenance Menus

To enter the maintenance menus press the

MENU/ENTER

switch.

If no switch is pressed for one minute, the

MX32 C

ONTROL

will return to normal display mode.

The request for Access Code display appears. The Access Code set at the factory is 1270.

Example of Displays

The first digit will
be flashing.

0000

cod

To enter the access
code. Press + or – to
select value of digit

1000

cod

Press ENTER to move
to next digit.

Press + or – to select
value of digit
Press ENTER to
move to next digit

1200

cod

Press + or – to select
value of digit

Press ENTER to move to
next digit

1270

cod

When access code is
complete,
press ENTER.

If the access code is invalid, the display will return to the normal mode.

If the access code is valid, the display will enter the maintenance menus

See Table 4 for the sequence of maintenance menus and a description of their uses. Press the + switch to move to the
next menu or the – switch to move back to the previous menu.

5.1.2 Exit the Maintenance Menus

To exit the maintenance menus at any point:

Press + & – switches at the same time. The EnrG display will appear.

Press + & – to go
to the save (exit)
display

enrg

oui

Press + or – to select OUI (yes) or NON (no). Press ENTER

Press + or – to
select yes or no

enrg

oui

The

MX32 C

ONTROL

will return to the normal display mode.

0