Enrg – ENMET MX-32 User Manual

Page 20

Advertising
background image

ENMET Corporation

MX32

E

NGUARD

16

15. Press + or – to select the stabilization time at start up. Press ENTER

Press + or – to
modify the time
digit by digit

1.00

min

Press ENTER when
correct time is selected

auT

man

Next the sensor type
will be flashing
May display:

OtH

16. Press + or – to scroll through the different types of sensor if necessary. Press ENTER

Press + or – to
modify the type
See choices below

AUT

man

Press ENTER when
correct sensor type is
selected

auT

man

Next the MAN or
AUT will be flashing

or

AUT

= Other, for toxic or Oxygen (for 4-20mA Sensor/Transmitter)

or

InC

= ionic, optical etc…

(not used)

EHP

or

EHP

= Combustible

(not used)

or

POnt

= Bridge

(not used)

17. Press + or – to select MAN or AUT. Press ENTER

MAN: no visual alarm, yellow LED flashing, when s/t is in calibration mode
AUT: visual alarm, yellow LED flashing, when s/t is in calibration mode

Press + or – to
select MAN or
AUT

AUT

man

Press ENTER when
correct choice is selected

0.120

min

Will be displayed

Press ENTER 4 times

enrg

non

Next the save (update)
programming will be
flashing

18. Press + or – to select OUI (yes) or NON (no). To update the programming for the channel, Press ENTER

If you select NON and press ENTER the programming will not be updated.
If you select OUI and press ENTER the programming will be updated.

Press + or – to
select yes or no

enrg

oui

Press ENTER when
correct choice is selected

0

co

Next the MX32
Control returns to the
normal display mode.

19. Programming is complete the

MX32

CONTROL

display returns to normal display mode.