Description of Event Types

354

MAX-700

Merged Event

Not All Event Types Are Tested

In this situation, the only thing that we can clearly know is when the loss
has a Pass status. If the global event loss is less than or equal to the
smallest threshold value (a value that is tested, of course), we are sure that
the merged event status is Pass. Otherwise, we cannot know, so the status
of the event is Unknown.

In our example, if we suppose that you chose not to test non-reflective
losses, then the analysis would be done as shown below:

Pass

Unknown

0

0.5

1.2

Pass

Unknown