Refer to the following table if your Glassware Washer fails to operate properly. If
the suggested corrective actions do not solve your problem, contact Labconco
Product Service for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Detergent cup in
door fails to open

Malfunction of cup

Replace cup.

Malfunction of relay

Check electrical continuity or replace relay PCB.

Glassware interfering
with cup door

Move glassware.

Liquid detergent
dispense > 0 ml

None

Noises during
operation

Glassware rattling

Load glassware properly. Do not allow glassware
to touch.

Door panel vibrating

Tighten panel screws.

Wash arms hitting

Check glassware loading.

Water inlet line
makes occasional
hammering sound

Check with plumber.

Purified water pump
is activated

No corrective action needed. This is normal for
about 5 minutes while pure water enters the
washer.

Door drops too far
(below door stop)

Bent door hinge

Replace door hinge.

Advertising