Labconco FlaskScrubber Vantage Series Glassware Washers 4540031 User Manual

Page 58

Advertising
background image

Chapter 7: Troubleshooting

Product Service 1-800-522-7658

54

54

PROBLEM

CAUSE

CORRECTIVE ACTION

Water Low Alarm

Improper water
supply

Make sure water is turned on.

Tap water pressure must be at least 20 psi
(138 kPa) and flow must be at least 1.25 gallons
(4.7 liters)/minute.

DI water supply must have flow of at least 0.9
gallons (3.4 liter)/minute.

At least 3.4 gallons (13 liters) must be available
for each DI rinse.

Defective fill valves

Clean screens inside the inlet fittings of the water
valves.

Check voltage to valves.

Replace valve if it does not operate.

Restricted flow

Check for kinks in water hoses in base of washer
to water inlet fitting on side of tank.

Defective DI water
pump

Check voltage at pump.

Listen for pump operation. Replace if necessary.

Defective hose to fill
switch

Check for splits.

Check for loose clamps.

Defective fill level
switch

Recalibrate switch. Contact Labconco Product
Service.

Replace switch.

Defective controller

Check output from PCB to both valves and DI
pump.

Replace PCB. Contact Labconco Product
Service.

Holes in filter screen
too small

Contact Labconco Product Service.

Defective grommet
where hose connects
to fill level switch

Replace grommet.

Advertising