Chapter 7: Troubleshooting

Product Service 1-800-522-7658

55

PROBLEM

CAUSE

CORRECTIVE ACTION

Low Temp Alarm

Heater inoperative

Check electrical connections.

Check voltage to heater.

Check resistance of heater. If open circuit,
replace heater.

Defective controller

Check output from PCB to heater.

Replace PCB. Contact Labconco Product Service.

Defective
temperature probe

Check electrical connections.

Contact Labconco Product Service. Replace if
necessary.

Defective high
temperature cut out
switch

Check resistance of safety cut out switch. If open
circuit at room temperature, replace switch.