C. set timer, D. set constant-potential charge mode, E. switch on mains power – Lamar Technologies CA 1550 User Manual

Page 28: F. connect external voltmeter

Advertising
background image

5. CALIBRATION AND MAINTENANCE

C. SET TIMER

Set the timer by pushing the up/down buttons to

a time that will allow calibration to be

completed before the time runs out (for

example 60 minutes).

D. SET CONSTANT-POTENTIAL

CHARGE MODE

Turn the Mode Select knob to Constant

potential, CP.

E. SWITCH ON MAINS POWER

Turn on the AC on-off/reset power switch.

The voltmeter reads the set voltage, the

ammeter reads zero. The timer reads 0’s and its

red LED is off. The Charge Mode Lamp

indicator is illuminated.

F. CONNECT EXTERNAL

VOLTMETER

Set the voltmeter to a DC voltage range

appropriate for the set constant potential.

Connect the external digital voltmeter into the

voltmeter calibration points.

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2