5. CALIBRATION AND MAINTENANCE

EXTERNAL CALIBRATED MILLIVOLTMETER

Figure 5-3. Circuit diagram for shunt verification

5.5 DISCHARGE CURRENT OPERATING RANGE

The CA-1550’s factory preset max discharge rate of amperes can be increased to 50

amperes. If necessary to increase the discharge rate please contact the factory.

5.6 MAINTENANCE

Standard electrical equipment maintenance and cleaning procedures should be

followed.

5.6.1 VENTS AND FAN

Regularly check that the fan vents next to the front panel is clean to ensure

adequate cooling of the unit. This is especially important when the unit is placed in a

dusty or otherwise dirty air environment.

5.6.2 DC BATTERY CABLE AND LEADS

Inspect DC battery cable and connector periodically. Replace damaged or worn

cable.

WARNING: Disconnect AC voltage and battery before

attempting to replace or secure cable.

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2