ROHO SelectAir MA mattress User Manual

Page 17

Advertising
background image

Filter

The filter should be inspected and cleaned every month under normal conditions.

Inspect filter and clean or replace as necessary. Bariatric patients generally require

higher air pressure for support, therefore, the air filter should be inspected and

replaced more frequently.

To check and change the filter:

. Firmly pull forward and then down on the filter

tray located at the front of the control unit.

The filter tray will automatically stop in the fully

opened position (fig. 11). It is not necessary to

completely remove the filter tray.

Important:

Keep the filter clean and the air

opening clear of obstacles.

2. Remove the filter from the filter tray (Fig. 12)

by grabbing the filter by its edges and

inspect the condition of the filter.

If dirty, wash the filter in a mild detergent.

Let the filter dry completely before

replacing the filter back into the filter tray.

If the filter is torn or in bad condition,

replace it with a new one.

Note: Do not use the control unit without

a filter.

3. Replace the filter by firmly pressing the filter

in around the edges in the filter tray.

4. Ensure the filter is laying flat and push the filter

tray up into the housing.

HealFloat Feature (HybridSelect only)
To properly maintain the ROHO HIGH PROFILE Cushion in the HealFloat feature,

refer to the cushion's operation manual.

Section 7

Maintenance

Figure 

Filter Tray

and Filter

Filter

Tray

Filter

Figure 

5